HF X-RAY MONOBLOCK FOR MOBILE APPLICATIONS

TECHNICAL SPECIFICATIONS

PARAMETER SPECIFICATION
Input Voltage 230V ±10% AC, 50Hz, single phase
Maximum Output Power
XRM3.5(3.5kW) XRM5(5kW) XRM6(6kW)
30mA at 100kV 50mA at 100kV 60mA at 100kV
80mA at 40kV 100mA at 50kV 150mA at 40kV
HF Generator Frequency 100kHz
Protections Over Voltage, Over Current, Over Temperature, Filament Fault, Earth Fault, Inverter Fault.
Application Intensive care units and radiology department
kV Range 40kV to 100kV (60 steps)
mA Range 20mA to 150mA (Depends on Monoblock power rating)
Modes APR & Manual
mAs 0.1 to 200mAs in 31 steps
Exposure Time 10ms to 5sec
Tube Compatibility XD3A
User Interface Keyboard with microcontroller-based control
Housing Monoblock
HF Inverter
Contains high voltage transformer, X-ray tube & filament transformers floating on high voltage
Contains inverter, control cards, filament card, transformers and other interlocks. (No external component required).
Size (L x W x H) Monoblock
HF Inverter
307mm x 149mm x 211mm
238mm x 238mm x 220mm
Weight Monoblock
HF Inverter
17 Kgs
10 Kgs