REGULATED HIGH VOLTAGE POWER SUPPLY FOR SPUTTERING

1kV DC, 1kW / 2kW / 5kW

TECHNICAL SPECIFICATIONS

PARAMETER SPECIFICATION
Input Voltage 230V ±10% AC, 50Hz, single phase / 415V ±10% AC, 50Hz, three phase
Output Voltage Range* 1kV DC
Output Power Range* 1kW / 2kW / 5kW
Polarity* Positive / Negative
Line Regulation Upto 0.05% for ±10% variation in input voltage
Load Regulation Upto 0.1% for 0 to 100% load variation
Stability < 0.01%/hour after 1 hour warm up
Temperature Coefficient <0.1%/ºC
Regulating Mode* Constant voltage – constant current
Voltage & Current Control By 10-turn potentiometers on the front panel from zero to maximum output rating
Remote Controls & Signals through Pluggable Connector (analog signals) / D-Connector (External RS232 Microcontroller Module for Remote Operation) 10V DC reference
10V DC HV enable signal
0 to 10V DC signal for voltage & current control
0 to 10V DC signal for voltage & current monitoring
Cooling Forced air
Protections* Against over load, arc and short-circuit
Front Panel Power ON/OFF switch / MCB with indication
HV ON/OFF switch with indication
3½ digit voltage and current meters
10-turn potentiometers for voltage and current control
Constant voltage – constant current mode indication
Back Panel Socket for mains input with 250V, 20A 3 conductor power cord of 1.5 meters / Terminal block for 3 phase inputs
Fuse holder with fuse
Terminal for HV output with 10ft of detachable high voltage cable
Stud for grounding the unit
Remote interface connector#
Topology High frequency resonant / PWM-controlled switch mode
Switching Device IGBT
Cabinet 19” rack, powder coated

*Optional. To be specified by the user.

Remote Interface Connector Configuration

Pins 14 Pin Pluggable Connector (Analog Interface) 25 Pin D Connector (PC Interface)
1 Earth Earth
2 Common Common
3 Interlock HV Enable
4 HV Enable Voltage Control Remote
5 Reference Current Control Remote
6 Common Over Voltage Control Remote*
7 Voltage Control Remote Over Current Control Remote*
8 Voltage Control Local Common
9 Common Voltage Monitor
10 Current Control Remote Current Monitor
11 Current Control Local Reserved Monitor*
12 Common Reserved Monitor*
13 Voltage Monitor Common
14 Current Monitor +Mode Status
15 Interlock Status
16 HV ON Status
17 Common
18 PS Fault Status*
19 Over Voltage Status*
20 Over Current Status*
21 Over Temperature Status*
22 Open Circuit Status*
23 Phase Failure Status*
24 Reserved Status*
25 Common