PARAMETER |
SPECIFICATION |
Input Voltage |
440V±5%, 50Hz, 3phase |
Radiography kV Range |
40kV DC to 125kV DC |
Radiography mA Range |
25mA to 500mA |
Radiography mAs Range |
0.25 to 500mAs |
Fluoroscopy kV Range |
40kV DC to 100kV DC |
Fluoroscopy mA Range |
0.5 to 4mA |
Fluoroscopy Time |
4min ON 8min OFF |
Maximum Output Power |
XR15D |
XR20D |
XR32D |
XR40D |
XR50D |
300mA at 50kV |
320mA at 62kV |
400mA at 80kV |
500mA at 80kV |
800mA at 100kV |
150mA at 100kV |
200mA at 100kV |
320mA at 100kV |
400mA at 100kV |
500mA at 100kV |
120mA at 125kV |
160mA at 125kV |
250mA at 125kV |
320mA at 125kV |
400mA at 125kV |
|
Application |
X-Ray Tube1: Radiography
X-Ray Tube2: Radiography & Fluoroscopy |
HF Generator Frequency |
40kHz |
Protections |
Over Voltage, Over Current, Over Temperature, Earth Fault, Filament Fault, Rotor Fault and Single Phasing. |
AEC |
Optional |
Tube Compatibility |
Toshiba E7239 / E7242 / (E7252 for 50kW) |
Remote Control |
Touch Screen tablet with microcontroller-based control / Keyboard |
Housing & Size |
MS cabinet without nylon wheels (616mm x 516mm x 810mm) (LxWxH) |