PARAMETER |
SPECIFICATION |
Input Voltage |
230V ±10% AC, 50Hz, single phase |
Maximum Output Voltage and Power |
50mA at 120kV 150mA at 40kV
|
HF Generator Frequency |
100kHz |
Protections |
Over Voltage, Over Current, Over Temperature, Filament Fault, Earth Fault, Rotor Fault, Inverter Fault. |
Application |
Intensive care units and radiology department |
kV Range |
40kV to 120kV (80 steps) |
mA Range |
20mA to 150mA (8 steps) |
Modes |
APR & Manual |
mAs |
0.1 to 200mAs in 31 steps |
Exposure Time |
10ms to 5sec |
User Interface |
Keyboard or Touch screen tablet with microcontroller-based control |
Housing |
Monoblock |
|
HF Inverter |
|
Contains high voltage transformer, X-ray tube & filament transformers floating on high voltage. |
Contains inverter, control & filament cards, transformers and other interlocks. (Rotor circuitry supplied separately). |
|
Size (L x W x H) |
Monoblock |
|
HF Inverter |
|
307mm x 149mm x 263mm |
238mm x 238mm x 220mm |
|
Weight |
Monoblock |
|
HF Inverter |
|
|